13
Calibration
CHARGE INJECTION CALIBRATION (CIC)
Injects an internally generated electrical signal in parallel with the
microphone diaphragm. A manual CIC can be performed whenever there
is no measurement in progress
An automatic CIC can be performed at the start and end of a logging
measurement
Storage
Measurement data is stored on an external memory device.
For availability, please refer to the Ordering Information
Notifications
ALARM CONDITIONS
CIC failed (in addition to those specified for BZ-7130)
Weather Data
MM-0256-A WEATHER STATION KIT
Six parameters: Wind Speed, Wind Direction, Ambient Temperature,
Ambient Humidity, Ambient Pressure, Rain Gauge)
MM-0256-A WEATHER STATION KIT
Two parameters: Wind Speed, Wind Direction
Software Specifications – Signal Recording Option BZ-7226
Signal Recording Option BZ-7226 is enabled with a separate license. It
works with all the software for Type 2250-L: Sound Level Meter,
Frequency Analysis and Logging Software
For data storage, Signal Recording requires:
SD Card
USB Memory Stick
RECORDED SIGNAL
A-, B-, C- or Z-weighted signal from the measurement transducer
AUTOMATIC GAIN CONTROL
The average level of the signal is kept within a 40 dB range, or the gain
can be fixed
SAMPLING RATE AND PRE-RECORDING
The signal is buffered for the pre-recording of the signal. This allows the
beginning of events to be recorded even if they are only detected later.
PLAYBACK
Playback of signal recordings can be listened to using the
earphone/headphones connected to the headphone socket
RECORDING FORMAT
The recording format is either 24- or 16-bit wave files (extension .WAV)
attached to the data in the project, easily played-back afterwards on a PC
using BZ-5503, Type 7820 or 7825. Calibration information is stored in
the .WAV file, allowing PULSE to analyse the recordings
Functions with BZ-7130 and BZ-7131
Manual Control of Recording: Recording can be manually started and
stopped during a measurement using a pushbutton or an external signal
Automatic Control of Recording: Start of recording when measurement
is started. Minimum and Maximum recording time can be preset
Functions with BZ-7133
Manual Control of Recording (using Manual Event or Back-erase
pushbutton, or an external signal): Recording during all of the event,
or for preset minimum and maximum duration. A Sound marker is set
while recording. Selectable pre- and post-recording time
Manual Control of Recording (using touch screen): Recording for the
selected time period (subject to the limitations of the pre-recording
buffer). A Sound marker is set for the selected time period
Automatic Control of Recording: Recording during all of the event or
for preset minimum and maximum duration. Selectable pre- and post-
recording time
Playback: Playback of signal recordings can be listened to using
earphone/headphones connected to the headphone socket
Gain Adjustment: –60 dB to 60 dB
Software Specifications – Tone Assessment Option BZ-7231
LICENSE
Tone Assessment Option BZ-7231 is enabled with a separate license and
can be used with the 1/3-octave and logging template (BZ-7132 and
BZ-7133)
TONE ASSESSMENT
Tone assessment is based on the measured 1/3-octave spectrum in
accordance with either the international ‘ISO 1996:2007 Acoustics –
Description, assessment and measurement of environmental noise – part
2: Determination of environmental noise levels. Annex D (informative)
Objective method for assessing the audibility of tones in noise –
Simplified method’
SPECTRA ASSESSED
The displayed 1/3-octave spectrum (L
eq
)
may be assessed. Assessment is made as post-processing, that is, when
measurement is paused or stopped
SETUP ACCORDING TO STANDARD
Setups in violation of the standard are indicated as such on the display.
You can then accept to apply the default setup. Tone assessment will be
made if possible, in spite of standard violations. For tone assessment
according to ISO 1996-2, Annex D, you can set the division between the
Low and Middle frequency range, the division between the Middle and
High frequency range, and the limits for the level differences between
adjacent bands
RESULTS
Tones are indicated above the spectrum when Tone is selected as
spectrum parameter. The resulting adjustment can be viewed on the
Value panel. It is not saved with the measurement
QUALITY INDICATORS
On the display, a quality indicator (smiley) will indicate that a hint is
available for tone assessment quality.
Click on the indicator to see the hint
Sampling
Rate
(kHz)
Maximum
Pre-recording (s)
16-bit
Maximum
Pre-recording (s)
24-bit
Memory
(KB/s)
16-bit
Memory
(KB/s)
24-bit
8 470 310 16 24
16 230 150 32 48
24 150 96 48 72
48 70 43 96 144