Hand-held Analyzer 2250 Light – User Manual
8.7 Timers and Automatic Measurements
Automatic control of the analyzer is available via one or more of the analyzer’s timers (up to
ten).
The purpose of a timer is to initiate measurements in the absence of an operator. Basically, a
timer will:
Switch on the analyzer at a preset time
Load a preselected template
Start a measurement as specified in the template
Stop the measurement after a preset time defined by the Timer Setup
Save the measurement
Switch the analyzer off
Repeat the above a preset number of times at preset intervals
For a timer to work, the analyzer must be in the ‘switched off’ state.
A timer will never interrupt the analyzer if it is already in use manually. If a switch-on time is
during the time the analyzer is in use, the timer opportunity is lost.
You can control the analyzer during a timer-initiated measurement as long as this does not
affect the measurements. If you try to change any parameters affecting the measurement you
will be warned of this. If you persist, the analyzer will revert to manual control and timer
control cannot be resumed.
The Timer Setup lists the timers that have been defined previously on the analyzer (i.e., added
by you or previous users).
Fig.8.8
Example Timer Setup