Transmitter Cond Ind 7100e/2(X)
Configuration
Current output
Concentration measurement: Select process solution.
Select sensor
Select measured variable
Enter current start
Enter current end
Set output filter
1. Press conf key.
2. Enter passcode 1200
*
.
3. Select Current output menu group
using arrow keys.
All items of this menu group
are indicated by the code “o1.”
4. Press enter to select menu,
edit with arrow keys (see Pg 43).
Confirm (and proceed) with enter.
5. End: Press conf, then enter
2
3
4
Current output:
22 mA in the case of error
Signal behavior during HOLD
1
5
Select solution (Conc)
Select characteristic (LIN/LOG)
* Factory setting