B&K Reference 7250 Series II Bedienerhandbuch Seite 45

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 108
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 44
45
Code
LIN
(LIN / LOG)
Display
Action
Choices
o1.
Select output characteristic
Select with arrow key
Proceed with enter
(Step omitted for % (Conc) or
SAL)
000.0 mS
(xxx.x mS)
100.0 mS
(xxx.x mS)
Conductivity
[mS/cm]
204
200
0
Assignment of measured values:
Current start and current end
[mS/cm]
204
200
100
[mA]
Example 1: Range 0...200 mS/cm
Example 2: Range 100...200 mS/cm
Advantage: Higher resolution in range
of interest
[mA]
Output current
Conductivity
With LIN selected:
Enter current start
(lower end of scale).
Select with key, edit number
with key, proceed with enter.
Enter current end
(upper end of scale).
Proceed with enter
Output current
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 45
Seitenansicht 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 107 108

Kommentare zu diesen Handbüchern

Keine Kommentare