88
Hand-held Analyzer Types 2250 and 2270 – User Manual
Table 8.2 Content of Notification
Note: Some of the information listed above may not be available, depending on the conditions
when the notification was sent, for example, no measurement values will be available
immediately after a power on.
8.4.2 Text/SMS Notifications
Notifications via SMS (text message) can be used if you have connected a
GSM/GPRS/EDGE/HSPA modem to the analyzer (see section 8.2.2) and set Modem to
Disabled (a text message cannot be sent if the modem is set to GPRS/EDGE/HSPA Dialup).
In addition to sending notifications initiated by alarms, you can send an ‘Info’ text to the
analyzer and ask for an SMS notification, after which you will get a notification within 30
seconds.
You need to setup a Phone Number in Preferences > Notifications > SMS and set Active to Yes
for the SMS. You can also add a Name for the number for easier identification.
You can setup two individual SMS phone numbers for sending the SMS notification to two
recipients.
Set Notification to SMS and a notification will be sent by SMS to the active phone numbers,
when any of the alarm conditions are met, or if the analyzer receives an ‘Info’ SMS.
Information Type Content
Reason for notification ‘Test’, ‘Alarm’ or ‘Status’ followed by a short description
Time of notification Date and Time
Instrument ID Serial No. and Location
Status of current Disk Free memory / Total memory
Status of Internal Battery Full, High, Low, Critical, Charging or Unknown
Trigger Input Voltage (if available) The voltage
Measurement Status (if available) Running, Paused or Stopped
No. of Signal Recordings (if
available)
The number of signal recordings made since start of
measurement
Latest CIC Result (if available) Passed, Background noise too high, Ratio outside tolerances,
Ratio deviates from Ref., Undefined Ref., or Undefined. For
2-channel measurements: One CIC result per channel
Current value of LAF (if available) The value. For 2-channel measurements: One CIC result per
channel
Report Threshold Level The value
Report Level LAeq The value