Hand-held Analyzer Types 2250 and 2270 User Manual
A.1 Input
Table A.1 Input parameters
Parameter Values Comment
Input Channel Ch. 1
Ch. 2
Determines whether Input Ch. 1 or Input Ch. 2 is
used for the measurement.
(Available in Type 2270 1-channel templates only)
Loudness
*
Free-field
Diffuse-field
Auto
Determines whether Loudness and Loudness Level
are calculated based on Free-field or Diffuse-field
conditions. Set to Auto to follow the setting of
Sound Field Correction. For an unknown
microphone type, you can set Loudness to the
condition matching your request.
Available for microphone input only.
Trigger Input
None
MATRON
Handswitch
Voltage Level
Voltage for
Monitoring
This parameter should be set to match the
equipment connected to the Trigger Input socket on
the connector panel of the analyzer.
Set it to None, if not used.
Set Trigger Input to MATRON Handswitch if the
analyzer is being used in the MATRON system.
This setting should also be used if you want to use
the Handswitch ZH-0680 for manually triggering
signal recordings.
Note: This can only be used together with
analyzers with serial numbers above 2479652
Set Trigger Input to Voltage Level if you want to
control signal recording by a voltage level
generated by external equipment. The Voltage
Level should generate at least 2 V for On and less
than 1 V for Off. The duration of the steady level
should be at least 1 s, so it can be recognised by
the analyzer
Set Trigger Input to Voltage for Monitoring if you
want to monitor the voltage at this input. This setting
can be used together with Notifications – see
section 8.4. This setting cannot be combined with
MATRON Handswitch or Voltage Level
Label Up to four
characters
Labels for the channels:
Leftmost is for Channel 1 (default set to "Ch.1")
Rightmost for Channel 2 (default set to "Ch.2").
(Available in 2270 2-channel templates only)