B&K 2250 Light Bedienungsanleitung Seite 187

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 332
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 186
CHAPTER 15
Tone Assessment Option BZ-7231 – 1/3-octave Method
175
15.5 Measuring
15.5.1 Result Display
Tone assessment is not performed until the Tone parameter is selected above the spectrum.
(This is done by tapping one of the parameter fields in the two lines above the Spectrum
display and choosing Tone from the drop-down menu.) Once selected, tone assessment is
carried out on the selected spectrum. The results are updated during the measurement.
Fig.15.8
Selecting the Tone
parameter above the
spectrum display
On the Results display, a blue marker is set at the top of each detected band with a tone.
Differences to the left and right of the frequency band selected by the main cursor are shown
(on the left and right) in the tone parameter panel above the spectrum (Fig.15.9). The main
cursor is moved using either the hard keys or the stylus.
Fig.15.9
The differences to the left
and right of the selected
frequency band are
shown in the tone
parameter panel – in this
example, there is a
11.4 dB difference to the
left and a 3.9 dB
difference to the right
Seitenansicht 186
1 2 ... 182 183 184 185 186 187 188 189 190 191 192 ... 331 332

Kommentare zu diesen Handbüchern

Keine Kommentare