B&K 2250 Light Bedienungsanleitung Seite 203

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 332
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 202
CHAPTER 17
Low Frequency Option BZ-7234
191
2) Select the Sound Level Meter, Frequency Analyzer, Logging, Logging SLM, Enh. Logging
or Enh. Logging SLM project template.
The project template is displayed at the top of the screen, if it does not display the correct
template tap the Project Template Bar at the top of the screen and select the correct
template from the drop-down menu. (See section 3.3.1 for more details on templates.)
3) Tap the Main Menu icon and select Setup.
4) Select Input > Transducer and select appropriate microphone.
5) Select Frequency Settings > Low Frequency Option and select On.
This will enable extra options for the Low Frequency parameter (Frequency Analyzer
module only) and for the frequency weighting parameters (Fig.17.2).
Fig.17.2
Example of additional
parameters with the
Low Frequency Option
parameter set to On
6) Set Low Frequency to Extended.
7) Set:
Broadband (excl. Peak) to A,G (simultaneous measurement with the two frequency
weightings A and G)
Broadband Peak to G if you need to measure the G-weighted peak
8) (Frequency Analyzer module only) Set Spectrum Weighting parameter to A, C, Z or G, as
required.
9) Tap to exit Setup.
Seitenansicht 202
1 2 ... 198 199 200 201 202 203 204 205 206 207 208 ... 331 332

Kommentare zu diesen Handbüchern

Keine Kommentare