Hand-held Analyzer Types 2250 and 2270 – User Manual
17.2.1 Controlling the Measurement
The measurement is controlled in the same way you would control a normal Sound Level
Meter measurement, using Start/Pause , Continue , Back-erase and Reset
pushbuttons, see Chapter 3 for more details.
Note 1: The low-frequency settings require much longer settling times of filters and detectors
for correct results. The instrument automatically takes care of this and waits (after you press
Start ) for the needed settling time for the individual detectors.
Note 2: You cannot start a measurement with G-weighting without using one of the low-
frequency microphones.
17.2.2 Displaying the Results
Tap on the broadband result readouts to select parameters from the Low Frequency or the
Lpeak group to select the G-weighted results (Fig.17.3).
Fig.17.3
Selection of Low
Frequency broadband-
readout results
Note: If you use the Frequency Analyzer module, you will also have the parameter
LXeq(f1-f2), X=A/B/C/Z/G and F1= bottom freq. and F2=top freq. of the frequency bands in
the spectrum. By using G-weighting of the spectrum you would, for example, be able to
readout the parameter LGeq(5 Hz-20 Hz) in cases where that needs to be reported.