CHAPTER 17
Low Frequency Option BZ-7234
193
17.3 Building Vibration Measurements
This section describes how to measure Building Vibration parameters. Building Vibration
parameters are measured using Wm or Wmb filters (
Fig.17.4
Wm filters
Low Frequency Option BZ-7234, together with BZ-7222, BZ-7223, BZ-7224 or BZ-7225 and
Accelerometer Type 8344 for low-level measurements, measures the following timed
broadband parameters in addition to the standard measured parameters:
aW – time-averaged weighted
*
acceleration value
PeakW – maximum peak of the weighted
*
acceleration value
aWCrest – crest factor, given by Peak
W
/a
W
MTVV – maximum of the a
W,1s
values
KBFmax – maximum of the KBF values
KBFTm – average of the values of KBFmax
i
, where each KBFmax
i
has been measured
during 30 s
The following instantaneous values are also available:
aW,1s – maximum time-weighted acceleration value during the latest second with
weighting1) and 1 s exponential time constant
KBF – instantaneous time-averaged acceleration level with W
m
weighting and with
125 ms exponential time constant
Note: W
m
weighting of the acceleration signal corresponds to using H
KB
weighting on the
velocity signal, as defined in DIN 45669–1
*. Weightings can be either W
m
(whole-body vibration in buildings, all directions) as defined in
ISO 8041:2005 or W
mb
, which is only the band-limiting part of the Wm filter. This is set by the Broadband
(Excl. Peak) parameter.
–80.00
–70.00
–60.00
–50.00
–40.00
–30.00
–20.00
–10.00
0.00
10.00
Gain [dB]
0.0100 0.1000 1.0000 10.0000 100.0000 1000.0000 10000.0000
Frequency [Hz]
W
mb
-Weighting W
m
-Weighting
130806