Hand-held Analyzer Types 2250 and 2270 User Manual
Level Difference Low 0.1 to 100.0 dB Set the required level difference for detecting
a tone in the low frequency range.
Default is 15 dB
Note: Only available if Tone Standard
parameter is set to ISO 1996:2-2007
Level Difference Middle 0.1 to 100.0 dB Set the required level difference for detecting
a tone in the middle frequency range.
Default is 8 dB
Note: Only available if Tone Standard
parameter is set to ISO 1996:2-2007
Level Difference High 0.1 to 100.0 dB Set the required level difference for detecting
a tone in the high frequency range.
Default is 5 dB
Note: Only available if Tone Standard
parameter is set to ISO 1996:2-2007
ISO 226 1987 Free-field
1987 Diffuse-field
2003 Free-field
The tone is checked against levels in other
frequency bands using normal equal loudness
level contours defined by ISO 226.
Select here which version of the standard to
use.
Note: Only available if Tone Standard
parameter is set to DM 16-03-1998
*. Requires Tone Assessment Option BZ-7231. Available for microphone input only.
Table A.19 (Cont.) Tone Assessment Parameters
*
Parameter Values Comment