Hand-held Analyzer Types 2250 and 2270 User Manual
C.2.2 Calibration History
Free-field Type Yes
No
Microphone only: Set to Yes for Free-field
types, otherwise set to No.
Set automatically for known Microphone type
Capacitance Double Microphone only: Insert capacitance of
microphone in pF.
Set automatically for known Microphone type
CCLD Yes
No
Set to Yes for CCLD (Constant Current Line
Drive) transducers, otherwise set to No.
Set automatically for known Microphone or
Accelerometer Type.
Note: Set Input to Rear Socket. The CCLD
input at the Rear socket will automatically be
enabled when selecting a transducer requiring
CCLD input
Calibration Check CIC
Actuator
Microphone only: For microphone
Type 4184-A you can select to check the
calibration of the system using CIC or
Actuator for all other microphone types the
setting is preset to CIC.
Weight Double Accelerometer only: Insert weight of the
accelerometer in g.
Set automatically for known Accelerometer
type
Table C.3 Calibration History parameters
Parameter Values Comment
1. Calibration date & Time YYYY-MM-DD hh:mm:ss Initial
1. Sensitivity Double Initial mV/Pa
1. Preamplifier ID No. Text string Initial
1. User Text string Initial
1. Input Top Socket, Rear Socket Initial
1. Calibration Type External,Internal Initial
1. Calibrator Serial No. Text string Initial
1. Comment Text string Initial
Table C.2 (Cont.) Transducer Setup parameters (for Microphone, Accelerometer or Direct, un-
less stated otherwise)
Parameter Values Comment