CHAPTER 8
Advanced Use of the Analyzer – Tips and Tricks
87
Table 8.1 Alarm Settings
Alarm Type Condition Comment
Power On When analyzer is switched
on or reboots
Indicates start from a timer, or a power failure, or
unexpected behaviour
Measurement
State
When a measurement
starts or stops
Indicates measurement finished
– or unexpected behaviour
CIC When Charge Injection
Calibration fails
Indicates possible measurement problem
Internal Battery When Internal Battery is
entering a specified state:
High, Low or Critical.
Indicates internal battery getting low. Internal Battery will
be in one of the states: Full, High, Low, Critical, Charging
or Unknown. You can set alarm to High, Low or Critical in
Preferences
> Notification Settings > Alarm
Settings
> Internal Battery.
Note: The internal battery will typically be in one of the
states Full or Charging if externally powered. Entering one
of the set states might then indicate a power failure
Trigger Input
Voltage
When voltage at Trigger
Input drops to or below a
specified level
Indicates no external power, or external battery getting low.
Set Setup
> Input > Trigger Type to Voltage for Monitoring.
Set the voltage limit in Preferences
> Notification
Settings
> Alarm Settings > Trigger Input Voltage.
Connect the external battery or external power to this input
(in addition to the External Power Socket) to get an alarm
if the voltage drops to or below the specified level
Disk Space When remaining disk
space drops to or below a
specified percentage of
the total disk space
Indicates disk space getting low
– needs to be swapped
with an empty disk. Set the percentage in
Preferences
> Notification Settings > Alarm Settings > Disk
Space
Level Trigger When level trigger
conditions are fulfilled
Indicates sound level exceeds trigger level specified in
Logging or Enh. Logging template
Report Level When L
Aeq
of the latest
Report Period exceeds
Report Threshold Level
Indicates that the sound level L
Aeq
for the latest Report
Period exceeds Report Threshold Level.
Note 1: If you specify a very low Report Threshold Level
you will receive a notification at the end of each Report
Period
Note 2: Only available when running the Enhanced
Logging template
Note 3: For 2-channel Enh. Logging the Report Threshold
Level is used for Ch.1 and Report Threshold Level 2 is
used for Ch.2.