CHAPTER 18
Specifications
209
CORRECTION FILTERS
For Microphone Types 4189, 4190, 4191, 4193, 4950,
4952, 4952+EH-2152, 4955-A, 4964 and 4184-A, BZ-
7227 is able to correct the frequency response to
compensate for sound field and accessories
Broadband Measurements
DETECTORS
A- and C-weighted broadband detectors with F
exponential time weighting
Overload Detector: Monitors the overload outputs of
all the frequency weighted channels
MEASUREMENTS
L
AF
and L
CF
for Display as Numbers or Quasi-
analogue Bars
MEASURING RANGES
When using Microphone Type 4189:
Dynamic Range: From typical noise floor to max.
level for a 1 kHz pure tone signal, A-weighted: 16.6 to
140 dB
Primary Indicator Range: In accordance with
IEC 60651, A-weighted: 23.5 dB to 122.3 dB
Linear Operating Range: In accordance with
IEC 61672:
• A-weighted: 23.5 dB to 122.3 dB
• C-weighted: 25.5 dB to 139.7 dB
• Z-weighted: 30.6 dB to 139.7 dB
Frequency Analysis
CENTRE FREQUENCIES
1/1-oct. Band Centre Frequencies: 63 Hz to 8 kHz
1/3-oct. Band Centre Frequencies: 50 Hz to 10 kHz
MEASUREMENTS
L
ZF
spectrum for display only
L
Zeq
spectra sampled at 5 ms intervals
MEASURING RANGES
When using Microphone Type 4189:
Dynamic Range: From typical noise floor to max.
level for a pure tone signal at 1 kHz 1/3-octave: 1.1 to
140 dB
Linear Operating Range: In accordance with
IEC 61260, 1/3-octave: 20.5 dB to 140 dB
Internal Generator
Built-in pseudo-random noise generator
Spectrum: Selectable Pink or White
Crest Factor:
• Pink noise: 4.4 (13 dB)
• White noise: 3.6 (11 dB)
Bandwidth: Follows measurement frequency range
• Lower Limit: 50 Hz (1/3-oct.) or 63 Hz (oct.)
• Upper Limit: 10 kHz (1/3-oct.) or 8 kHz (oct.)
Output Level: Independent of bandwidth
• Max.: 1 Vrms (0 dB)
• Gain Adjustment: –60 to 0 dB
When bandwidth is changed, the level for all bands is
automatically adjusted to comply with the set output
level
Correction Filters for Sound Sources Type 4292,
Type 4295 and Type 4296: Flat or Optimum
Turn-on Time and Turn-off Time: Equivalent to
RT = 70 ms
Repetition Period: 175 s
Output Connector: Output Socket
Control: See Measurement Control
External Generator
Selectable as alternative to internal generator
For controlling external noise generator
Levels: 0 V (Generator off), 4.5 V (Generator on)
Rise-time and Fall-time: 10 s
Control: See Measurement Control
Reverberation Time
EDT, T20 and T30 in octave or 1/3-octave bands
Decays: Measured and stored using averaging time of
5ms
Evaluation Range: 0 to –10 dB for EDT, –5 to –25 dB
for T20 and –5 to –35 dB for T30
Measurement Time: Automatic selection of
measurement time for the decays based on the actual
reverberation time of the room
Maximum Measurement Time: from 2 to 30 s
Averaging: EDT, T20 and T30 measurements can be
averaged (arithmetic averaging or ensemble
averaging)
EDT, T20 and T30 Calculation: From slope in
evaluation range
Slope Estimation: Least squares approximation
Quality Indicators: Quality Indicators with status
information like Overload, Curvature in %, etc.;
extensive list of status information.
Quality indicators are available on reverberation time
spectra for each frequency band, and as overall quality
indicators for each measurement position and for the
total project (room)
Reverberation Time Range: Max. 50 s, min.
0.1 – 0.7 s, depending on bandwidth and centre
frequency
Wide Band Reverberation Time: The arithmetic
average of the Reverberation Time within a selectable